EMF Exposure Systems
SAR slice banner
sXEs

sXEs


The sXEs system for exposure
of electrosensitive persons at
their homes is designed to have
a well defined exposure while at
the same time being highly portable. The ELF and RF systems
are based on transport casings
(< 18kg) and can be operated
by a tablet computer. Both
systems allow incident field
strengths up to the guideline
limits of ICNIRP for the general
public.

sXEs human exposure system
 

 Size

Exposure system (portable transport cases)

15 × 40 × 1200 cm (d × w × h)

 Signals

Carrier frequency

RF: 800/900/1800/1950/2400 MHz (broadband antenna)
ELF: up to 6 kHz

  Modulation schemes

RF: CW, 3G, 4G, DECT, GSM, WIFI (any I/Q modulation on request)
ELF: sine, powerline, “dirty currents”

  Intermittent exposure arbitrary on/off cycles with 10 s resolution

 Dosimetry

Incidence Field Level

RF: up to > 5 V/m
ELF: up to > 100uT

  Exposed area whole torso and head of the volunteer
  SAR characterization

only incident field is characterized. RF exposure is in the far-field of the antenna

 Exposure Control

· full software control of the exposure with self-detection of malfunctions
· single and double-blinded protocols
· radiated power monitoring and feedback regulation
· “pilot” exposure mode for pretesting
· data logging every 10s

Blinded protocols: single and double-blind protocols implemented on the control software. A set of exposure blocks can be guessed by the volunteers.

The complete bundle includes the exposure apparatus and PC with controlling software. Also included are personnel training sessions. Pricing and detailed specifications available on request.

For more information please contact us: customized@itis.swiss.