EMF Exposure Systems
SAR slice banner
sXc2450

sXc2450

 

Designed to test radiofrequency electromagnetic field exposures from a choice of second, third, fourth or fifth generation (2G, 3G, 4G, 5G) mobile communication devices or any combination of these operating at 2450MHz. The setup is based on two R22 rectangular waveguide chambers that are blindly excited by a fully computer-controlled signal unit.

sXc2450 web big
 

 Size

Exposure setup (placed inside incubator) 43 × 20 × 36 cm (d × w × h)
  Signal tower (beside incubator) 60 × 60 × 80 cm (d × w × h)

 Cell Cultures

Cell type any
  Cell preparation monolayer: 6 Petri dishes (NUNC,
ø = 35 mm)
  Loading volume of cell medium 18 mL

 Signals

Carrier frequency 2450 MHz
  Choice of Modulation schemes 5G
LTE (4G)
UMTS (3G)
GSM nonDTX
GSM DTX
GSM talk
arbitrary pulse structure
continuous wave
any AM waveform of 64000
points on request
WiFi Signal on request
  Intermittent exposure arbitrary on/off cycles with 10 s
resolution

 Dosimetric Key
 Parameters

Peak SAR (100 W amplifier) monolayer: 150 W/kg (DMEM)
  Nonuniformity of the SAR distribution monolayer: < 30%
  Temperature rise for exposed cells monolayer: 0.03°C/(W/kg)

 Exposure and
 Environmental Control

· full software control of the exposure with self-detection of malfunctions
· blinded exposure protocols
· incident field monitoring and feedback regulation by H-field loop sensors
· air temperature monitoring by Pt100 probes
· monitoring of fan driving currents
· data logging every 10 s
 

The complete bundle includes the waveguide chambers, signal tower, PC with software, and a suitable CO2 incubator. Also included is installation of the setup at the target laboratory as well as personnel training sessions. Pricing and detailed specifications available on request.