Designed to test radiofrequency electromagnetic field exposures from fifth generation (5G) mobile communication devices operating at 27.5GHz.
The setup is based on three rectangular chambers (High, Low, Sham) that are blindly excited by a fully computer-controlled signal unit.
Size |
Exposure setup (placed inside incubator) | 27 × 54 × 29 cm (d × w × h) |
Signal tower (beside incubator) | 60 × 60 × 80 cm (d × w × h) | |
Cell Cultures |
Cell type | any |
Cell preparation | 4x 25cm2 flasks | |
Loading volume of cell medium | min 7.5mL / flask | |
Signals |
Carrier frequency | 27.5 GHz |
Choice of Modulation schemes | 5G other signals on request |
|
Intermittent exposure | arbitrary on/off cycles with 10 s resolution | |
Dosimetric Key
|
SAR | 20 W/m2 average 500 W/m2 peak |
Nonuniformity of the SAR distribution | monolayer: ~30% | |
Temperature rise for exposed cells | 0.1°C for 6.3 W/m2 avg | |
Exposure and
|
· full software control of the exposure with self-detection of malfunctions · blinded exposure protocols · incident field monitoring and feedback regulation by H-field loop sensors · air temperature monitoring by Pt100 probes · monitoring of fan driving currents · data logging every 10 s |
The complete bundle includes the waveguide chambers, signal tower, PC with software, and a suitable CO2 incubator. Also included is installation of the setup at the target laboratory as well as personnel training sessions. Pricing and detailed specifications available on request.
For more information please contact us: customized@itis.swiss