EMF Exposure Systems
SAR slice banner
sXc_mm-wave

sXc mm-wave

 

Designed to test radiofrequency electromagnetic field exposures from fifth generation (5G) mobile communication devices operating at 27.5GHz.

The setup is based on three rectangular chambers (High, Low, Sham) that are blindly excited by a fully computer-controlled signal unit.

 

 Size

Exposure setup (placed inside incubator) 27 × 54 × 29 cm (d × w × h)
  Signal tower (beside incubator) 60 × 60 × 80 cm (d × w × h)

 Cell Cultures

Cell type any
  Cell preparation 4x 25cm2 flasks
  Loading volume of cell medium min 7.5mL / flask

 Signals

Carrier frequency 27.5 GHz
  Choice of Modulation schemes 5G
other signals on request
  Intermittent exposure arbitrary on/off cycles with 10 s resolution

 Dosimetric Key
 Parameters

SAR 20 W/m2 average 
500 W/m2 peak
  Nonuniformity of the SAR distribution monolayer: ~30%
  Temperature rise for exposed cells 0.1°C for 6.3 W/m2 avg

 Exposure and
 Environmental Control

· full software control of the exposure with self-detection of malfunctions
· blinded exposure protocols
· incident field monitoring and feedback regulation by H-field loop sensors
· air temperature monitoring by Pt100 probes
· monitoring of fan driving currents
· data logging every 10 s
 

The complete bundle includes the waveguide chambers, signal tower, PC with software, and a suitable CO2 incubator. Also included is installation of the setup at the target laboratory as well as personnel training sessions. Pricing and detailed specifications available on request.

For more information please contact us: customized@itis.swiss