The sXcLive2450 is a miniature RF exposure system at 2450 MHz, which is integrated with a microscope system. State-of- the-art microscopic techniques, including confocal scanning and fluorescence microscopy, allow live cell imaging during expo- sure. In contrast to the classic approach of post analysis of cell responses, the instrument enables direct insight into the cell’s response to EMF.
Size |
Exposure Element (placed on microscope) | 10 × 10 × 3 cm (d × w × h) |
Microscope/incubator | 60 × 60 × 80 cm (d × w × h) | |
Control device including computer | 40 × 40 × 10 cm (d × w × h) | |
Cell Cultures |
Cell type | cell monolayer |
Cell preparation | on plate in the live imaging module | |
Observation area | ø = 10 mm, 3.1 cm³ | |
Signals |
Carrier frequency | 2450 MHz |
Modulation schemes | WiFi modulation arbitrary pulse structure continuous wave any AM waveform of 64000 points on request |
|
Intermittent exposure | arbitrary on/off cycles with 10 s resolution | |
Dosimetric Key
|
Nonuniformity of the bottom layer | (3mm radius): 0.13dB |
Average SAR | >100 W/kg for 1 W input power | |
Exposure and
|
· full software control of the exposure with self-detection of malfunctions · sequentially blinded exposure protocols · current monitoring and feedback regulation · air temperature control and monitoring · monitoring of fan driving currents · data logging every 10 s |
The complete bundle includes the exposure cavity, signal tower and PC with controlling software. Also included is installation of the setup at the target laboratory as well as personnel training sessions. The live imaging microscope/incubator is not included. Pricing and detailed specifications available on request.
For more information please contact us: customized@itis.swiss.