EMF Exposure Systems
SAR slice banner
sXcLive2450

sXcLive2450

 

The sXcLive2450 is a miniature RF exposure system at 2450 MHz, which is integrated with a microscope system. State-of- the-art microscopic techniques, including confocal scanning and fluorescence microscopy, allow live cell imaging during expo- sure. In contrast to the classic approach of post analysis of cell responses, the instrument enables direct insight into the cell’s response to EMF.

sXcLive2450 in vitro exposure system
 

 Size

Exposure Element (placed on microscope) 10 × 10 × 3 cm (d × w × h)
  Microscope/incubator 60 × 60 × 80 cm (d × w × h)
  Control device including computer 40 × 40 × 10 cm (d × w × h)

 Cell Cultures

Cell type cell monolayer
  Cell preparation on plate in the live imaging module
  Observation area ø = 10 mm, 3.1 cm³

 Signals

Carrier frequency 2450 MHz
  Modulation schemes WiFi modulation
arbitrary pulse structure
continuous wave
any AM waveform of 64000 points on request 
  Intermittent exposure arbitrary on/off cycles with 10 s resolution

 Dosimetric Key
 Parameters

Nonuniformity of the bottom layer (3mm radius): 0.13dB
  Average SAR >100 W/kg for 1 W input power

 Exposure and
 Environmental Control

· full software control of the exposure with self-detection of malfunctions
· sequentially blinded exposure protocols
· current monitoring and feedback regulation
· air temperature control and monitoring
· monitoring of fan driving currents
· data logging every 10 s
 

The complete bundle includes the exposure cavity, signal tower and PC with controlling software. Also included is installation of the setup at the target laboratory as well as personnel training sessions. The live imaging microscope/incubator is not included. Pricing and detailed specifications available on request.

For more information please contact us: customized@itis.swiss.