The sXv mm-wave system is designed to expose 3 chambers (High, Low, Sham)of 12 cages each to fifth generation (5G) RF frequencies with a carrier frequency of 27.5 GHz. The reverberation chamber design allows unrestrained motion of the rodents inside their cages.
Size |
Exposure setup | 102 × 81 × 205 cm (d × w × h) |
Control device including computer | 60 × 60 × 80 cm (d × w × h) | |
Signals |
Carrier frequencies | 27.5 GHz |
Modulation schemes | 5G Other signals on request |
|
Intermittent exposure | arbitrary on/off cycles with 10 s temporal resolution | |
Exposure |
Mice |
12 cages Up to 4 mice / cage |
Dosimetric Key
|
Exposure isotropy | 0.86 dB |
Incident Power Density |
20 W/m2 avg ~500 W/m2 peak |
|
Exposure Control |
· full software control of the exposure with self-detection of malfunctions · immediate abortion in case of overexposure · blinded exposure protocols · air temperature and humidity monitoring · data logging every 10 s |
The complete bundle includes the exposure cavities, signal tower and PC with controlling software. Also included is installation of the setup at the target laboratory as well as personnel training sessions. Pricing and detailed specifications available on request.
For more information please contact us: customized@itis.swiss.