EMF Exposure Systems
SAR slice banner
sXv_mm-wave

sXv mm-wave

 

The sXv mm-wave system is designed to expose 3 chambers (High, Low, Sham)of 12 cages each to fifth generation (5G) RF frequencies with a carrier frequency of 27.5 GHz. The reverberation chamber design allows unrestrained motion of the rodents inside their cages.

 

 Size

Exposure setup 102 × 81 × 205 cm (d × w × h) 
  Control device including computer 60 × 60 × 80 cm (d × w × h)

 Signals

Carrier frequencies 27.5 GHz
  Modulation schemes 5G
Other signals on request
  Intermittent exposure arbitrary on/off cycles with 10 s temporal resolution

 Exposure

Mice

12 cages

Up to 4 mice / cage

 Dosimetric Key
 Parameters

Exposure isotropy 0.86 dB
     
  Incident Power Density

20 W/m2 avg

~500 W/m2 peak

 Exposure Control

· full software control of the exposure with self-detection of malfunctions
· immediate abortion in case of overexposure
· blinded exposure protocols
· air temperature and humidity monitoring
· data logging every 10 s
 

The complete bundle includes the exposure cavities, signal tower and PC with controlling software. Also included is installation of the setup at the target laboratory as well as personnel training sessions. Pricing and detailed specifications available on request.

For more information please contact us: customized@itis.swiss.