The sXv27 system is designed to expose 2 × 8 cages with 2 mice each to 27 MHz RF signals. The resonant TEM cell design allows unrestrained motion of the rodents inside their cages.
Size |
Exposure Element | 182 × 100 × 50 cm (d × w × h) resonator coaxial cable 2 × 181.5cm (including connectors) |
Control device including computer | 40 × 40 × 10 cm (d × w × h) | |
Signals |
Carrier frequency | 27 MHz |
Modulation schemes | 1 Hz – 200 kHz sinusoidal modulation (sequentially adjustable with 1 s resolution) continuous wave |
|
Intermittent exposure | arbitrary on/off cycles with 10 s temporal resolution | |
Exposure |
Mice | 2 mice × 8 cages × 2 chambers |
Dosimetric Key
|
E-field homogeneity | 0.45 dB |
Maximum field strength | 41 V/m for 1 W input power in the non-resonant case | |
SAR levels | >1 W/kg | |
Exposure Control |
· full software control of the exposure with self-detection of malfunctions · blinded exposure protocols · air temperature and humidity monitoring · data logging every 10 s |
The complete bundle includes the exposure cavities, signal tower and PC with controlling software. Also included is installation of the setup at the target laboratory as well as personnel training sessions. Pricing and detailed specifications available on request.
For more information please contact us: customized@itis.swiss.