EMF Exposure Systems
SAR slice banner
sXv27

sXv27

 

The sXv27 system is designed to expose 2 × 8 cages with 2 mice each to 27 MHz RF signals. The resonant TEM cell design allows unrestrained motion of the rodents inside their cages.

sXv27 in vivo exposure system
 

 Size

Exposure Element 182 × 100 × 50 cm (d × w × h)
resonator coaxial cable 2 × 181.5cm (including connectors)
  Control device including computer 40 × 40 × 10 cm (d × w × h)

 Signals

Carrier frequency 27 MHz
  Modulation schemes 1 Hz – 200 kHz sinusoidal modulation (sequentially adjustable with 1 s resolution)
continuous wave
  Intermittent exposure arbitrary on/off cycles with 10 s temporal resolution

 Exposure

Mice 2 mice × 8 cages × 2 chambers

 Dosimetric Key
 Parameters

E-field homogeneity 0.45 dB
  Maximum field strength 41 V/m for 1 W input power in the non-resonant case
  SAR levels >1 W/kg

 Exposure Control

· full software control of the exposure with self-detection of malfunctions
· blinded exposure protocols
· air temperature and humidity monitoring
· data logging every 10 s
 

The complete bundle includes the exposure cavities, signal tower and PC with controlling software. Also included is installation of the setup at the target laboratory as well as personnel training sessions. Pricing and detailed specifications available on request.

For more information please contact us: customized@itis.swiss.